OzoneLab Menu / Help
OzoneLab(TM) Instruments
Spacer ImageOzoneLab™ GE60 Ozone Generator Output Test Report Example Site SearchContact Us
Files ImageHome > Products > Products > Generators > Output & Graphs > GE30 Test Report Example

Please Note: We have discontinued our GE Product Line of Ozone Generators (GE30 & GE60). The GE60/FMxx has been replaced by the OL80W/FMxx which is a more energy efficient product with better performance and increased versatility (EXT port).

Flow in Liters per Minute (LPM) -> 1 LPM -> 1000 ml/min (1000 ml/min)
Ozone Concentration in Gamma -> 1 gamma -> 1µg/ml -> 1 mg/l -> 1gr/m3

Temp: 21 C
Feed gas: Oxygen/Med.
Pressure: 100.1 kPa
Note: YY/5.5/120/714/BT/15C/12VDC
Date: Aug 22, 1999
Model: GE60/12VDC
Serial #: 990822.b
File: 990822DC.OUT
Flow in Liters Per Minute (LPM).  Ozone concentration in GAMMA (µg/ml)

FLOW FLOW
------- OZONE CONCENTRATION REGULATOR SETTING ----------
LPM cc/min 1 2 3 4 5 6 7 8 9 10

1/32 31 22 29 38 46 53 59 68 77 84 90
1/16 62 10 13 19 24 29 34 44 53 63 73
1/8 125 7 9 13 16 20 24 32 39 50 60
1/4 250 4 5 7 9 12 14 19 23 30 36
1/2 500 3 4 5 5 6 7 10 12 16 19
3/4 750 2 3 4 4 5 5 7 9 11 12
1 1000 1 2 3 3 4 4 6 7 9 10

Test - concentration regulator setting #1, #2, #4, #6, #8 and #10 with oxygen flow from 1/32 LPM to 1 LPM (bold print).
Estimate - setting #3, #5, #7 and #9 (plain print)

Please see out page on How to Read the Ozone Test Report for an explanation of the above report.


© Ozone Services & OzoneLab™ For problems pertaining to our Web Site, please contact our WebMaster